Laboratory Equipment
Circular Rotary Vibrating Screen – Model: CQXZS-1000-2S
R 121,900.00View specifications

The S-20 and S-22 are high quality, low cost measurement systems for routine analysis in QC applications, education and training.
Catalogue price shown. Availability, delivery and the final total are confirmed by quotation. Online payments are not yet enabled.
UV/VIS SPECTROPHOTOMETER
Lead Time: 3-4 Weeks
The S-20 and S-22 are high quality, low cost measurement systems for routine analysis in QC applications, education and training.
| Specification | S-200 Vis | S-220 UV/Vis |
| Wavelength Range: | 320 to 1100nm | 190 to 1000nm |
| Spectral Bandwidth: | 6nm | 5nm |
| Transmittance accuracy: | ±0,5% T (NIST 930 Filter) | ±1% T (NIST 930 Filter) |
| Transmittance repeatability: | 0.2% T | 0.5% T |
| Baseline flatness: | ± 0.002 Abs (330-1090nm) | ± 0.005 Abs (200-990nm) |
| Noise level: | ? 0.001 Abs (500nm) | ? 0.005 Abs (250 nm) |
| Baseline stability: | ? 0.001 Abs/h (500nm) (after 2 hours warm up) | ? 0.005 Abs/h (250nm)(after 2 hours warm up) |
| Stray light: | ? 0.5% T | ? 0.5% T |
| Wavelength controlled variable: | 0.2nm | 0.2nm |
| Wavelength accuracy: | ± 1nm | ± 2nm |
| Wavelength repeatability: | 2400nm/min (0,2 sampling interval without filter) | 300nm/min (0,2 sampling interval without filter) |
| Wavelength scan speed: | (0,2 sampling interval without filter) | (0,2 sampling interval without filter) |
| Wavelength move speed: | to any specified position within 1sec. | to any specified position within 1sec. |
| Absorbance: | -0.3 to 1.999 | -0.3 to 1.999 |
| Transmittance: | 0 to 199.9% | 0 to 199.9% |
| Spectrum Scanning: | Yes | Yes |
| Concentration: | -300 to 1999 | -300 to 1999 |
| Selectable Resolution: | 1, 0.1, 0.01 or 0.001 | 1, 0.1, 0.01 or 0.001 |
| Light source: | Tungsten Halogen lamp | pulsed-Xenon lamp |
| Detector: | Silicon photodiode | Silicon photodiode |
| Display screen: | 4,3 inches colorful touch LCD screen | 4,3 inches colorful touch LCD screen |
| Printer: | specified 80-column thermal printer (series port) | specified 80-column thermal printer (series port) |
| Metering mode: | Single beam | Single beam |
| Memory: | SD card storage | SD card storage |
| Time Scan: | Graphical and calculated concentration value | Graphical and calculated concentration value |
| Analysis: | Absorbance and wavelength of peaks and valleys | Absorbance and wavelength of peaks and valleys |
| GLP: | Real time clock and calendar, Self Diagnosis | Real time clock and calendar, Self Diagnosis |
| Size: | 400 (W) x 280 (D) x 160 (H) mm | 400 (W) x 280 (D) x 160 (H) mm |
| Power requirement: | AC, 100-240V, 50/60Hz | AC, 100-240V, 50/60Hz |
| Power consumption: | 100VA | 100VA |
| Communication ports: | Serial printer port connects thermal printer, USB port connects PC, SD card port saves data and measurement methods, Accessories port connects and controls serval options | Serial printer port connects thermal printer, USB port connects PC, SD card port saves data and measurement methods, Accessories port connects and controls serval options |
| Weight: | 4 kgs | 4 kgs |